Next: Software
Up: No Title
Previous: The Siesta Datalogger
Subsections
Sensors
The sensors used in this project had the task of collecting information
relating to ambulation of a subject, which was then recorded and transmitted
to the software for processing.
A number of different sensors and configurations were trialled, predominately
concentrating on collecting information from the lower leg and foot. Two
primary types of sensors were used; accelerometers and pressure transducers.
Compumedics supplied some of their manufactured piezoelectric accelerometers,
designed to be used with the Siesta to detect movement during sleep
recordings. As such, these sensors were designed to be strong, cheap and
produce qualitative information relating to movement.
After examining the piezoelectric accelerometer response, it was decided that
some different sensors for producing quantitative information relating to foot
up or down state - pressure transducers - were required. This chapter
discusses the results of investigations into the piezoelectric sensors'
response, their applicability to ambulation monitoring, implementation of the
pressure transducers and results from other sensors which were briefly tested
but not used.
Piezoelectric accelerometers
The piezoelectric accelerometers supplied by Compumedics utilise the
piezoelectric properties of a polarised polymer. When the polymer is
disfigured, it produces a small voltage which is recorded by the Siesta.
The term `piezo' derives from the Greek `pressure' or `to press' and is
defined as ``polarisation in a substance resulting from application of
mechanical stress, especially in certain crystals'' [56].
The piezoelectric effect occurs in crystals such as quartz, which produce
small voltages across its faces when subjected to mechanical pressure
[13]. Conversely the piezoelectric effect works in reverse -
applying a voltage to a piezoelectric substance causes the crystal structure
to deform, resulting in movement.
Due to these effects, piezoelectric devices find wide use in electronics
applications such as high-frequency oscillators[57] and
transducers. Applications such as accelerometers, microphones and audio
speakers usually use a specially treated polymer film which produces voltage
between two points (or movement) when subjected to pressure (or a voltage).
Figure 6.1:
Compumedics supplied piezoelectric sensor and connecting leads
|
|
The response of the piezoelectric accelerometers was predominately AC (high
frequency), with little to no DC component. An example of the accelerometer
response is shown in Figure 6.2. This data was recorded
from sensors attached to the top of each shoe while taking two slow steps, and
shows peaks when the foot starts to swing and then strikes. It was discovered
that in practice it was very difficult to reliably distinguish between a
foot-strike and toe-off (Section 2.1.1), or extract distinct
events from the multiple signal peaks usually resulting from a single
movement. These sensors did provide useful evidence of fast and sharp
movements however, such as a hard footfall that may occur when correctional
steps are utilised to maintain balance.
Figure 6.2:
Example piezoelectric accelerometer response
|
|
Another method of using the piezoelectrics trialled was to place them under
the in-sole in the shoe to measure pressure as weight is applied and removed
from each foot (see Figure 6.3). The sensor
response tended to give large `spikes' when the load was applied and removed.
Again, this produced problems with distinguishing between foot up and down.
Figure 6.3:
Response of piezoelectric sensor placed in shoe in-sole
|
|
After much work with the piezoelectrics and attempts to extract concrete
ambulation information from the signals, it was concluded that the
piezoelectric sensors alone could not provide enough quantitative (and to an
extent, even qualitative) information for the project's purposes. Although not
particularly well suited for providing evidence relating to the subject's
stage in the gait cycle, or whether the feet were on or off the ground, they
were still considered useful for detecting the accelerations that may occur
during a stumble of fall. The high-frequency noise and interference of the
sensors could be handled using a combination of software filtering and an
`unreliable' sensor model in the DBN.
In the previous project by McGowan [4], force-sensitive resistors
(FSRs) were successfully used as pressure transducers, but it was discovered
similar sensors are very difficult to come by in Australia. Some investigation
lead to Associate Professor Andy Russell from the Department of Electrical and
Computer Systems Engineering, who through his work with tactile sensors for
robotics possesses a wide array of pressure-sensitive resistors and materials.
He provided samples of some FSRs and a sample of material called `Vermahide'.
The delicate nature of the manufactured FSRs resulted in the design and
implementation of some new pressure transducers using Vermahide.
McGowan noted in his thesis that the pre-packaged force-sensitive resistors
used in his project were liable to breakdown as a result of the large forces
exerted on them when used as foot sensors. Due to the increased likelihood of
failure in these sensors, compared to the relative robustness, flexibility
and adaptability of the Vermahide material, it was decided that pressure
transducers involving Vermahide would be more suitable as foot-sensors. As a
result a variety of sensors involving different configurations of Vermahide
and electrical contacts were built and tested.
Vermahide pressure transducers
Vermahide [58] is a felt-like fabric composed of openly-matted
synthetic fibres, which are encased in a conductive carbon emulsion. The
emulsion itself is not pressure-sensitive, but when pressure is applied to the
material, the fibres move closer together, and their effective conductive
paths from one side of the material to the other are reduced, resulting in a
drop in resistance. The matted fibre structure of the material gives it quite
a good elastic memory, in that it will tend to restore to its original shape
after pressure is removed.
Some initial sensors developed using aluminium contacts sandwiching a piece of
Vermahide; this worked well but it was found that the leads to the sensors did
not solder well to the Aluminium, and could easily be pulled off. Under the
advice of Professor Andy Russell, the use a of printed circuit board (PCB) was
trialled. The first Vermahide-PCB sensor used a pre-printed PCB with 2 mm wide
parallel copper tracks, separated by 1 mm. The two connecting leads for the
sensor were electrically connected to alternating copper tracks by way of
short soldered connecting leads. The Vermahide was sandwiched in between the
PCB and a stiff non-conducting material (cardboard was used) so that it
contacted the tracks. Hence if pressure was applied the resistance measurement
was only taken from one side of the Vermahide.
The alternating track connections required delicate wiring and soldering,
increasing the risk of the sensor becoming a short circuit, resulting in large
signals which may damage the Siesta.
In an attempt to overcome the above shortcomings, the final design use a
mm of Vermahide between two pieces of one-sided plain
copper-plated PCB. A small hole was drilled through the PCB so the lead could
be passed through from the un-plated the to plated side and soldered to the
copper which was in contact with the Vermahide. As in the aluminium-contact
sensors, the Vermahide was sandwiched between two pieces of PCB so that the
resistance across the material could be measured, as shown in Figure
6.4. The whole assembly and protruding leads were
secured with electrical tape to keep the sensor's thickness to a minimum.
This configuration was used for building more sensors, as it appeared the most
robust and simple implementation, reducing the risk of sensor failure due to
short circuits or disconnected leads. During testing, no such problems were
encountered with these sensors.
Figure 6.4:
Exploded view of Vermahide pressure sensor
|
|
Circuit design
As the Vermahide sensor changes resistance with pressure, but the Siesta
expects voltage inputs, a small circuit was required to produce a voltage
relative to the change in the sensor's resistance due to applied pressure.
This circuit was a voltage divider, consisting of power source VS (a 1.5
volt AA battery), a fixed resistor R and the foot sensor RF, which acts
like a variable resistor.
Figure 6.5:
Voltage divider circuit diagram, where `Rf' represents the
Vermahide sensors and `R1' to `R4' are the fixed biasing resistors.
|
|
As each of the four Vermahide-based sensors needed to give independent
signals, the circuit based on the diagram in Figure
6.5, consisting of four parallel voltage dividers (one
for each sensor), was implemented. The four fixed resistors, sensor leads and
connectors for the Siesta datalogger were soldered to a piece of pre-printed
circuit board and placed in a small containing box with the battery to prevent
damage.
Figure 6.6:
Four Vermahide pressure sensors, box containing voltage dividers,
`AA' battery, and connecting leads. Siesta shown with back cover
removed, exposing batteries and input channel connectors.
|
|
Given the voltage source value VS, the resistance values of R and the
sensor RF, the output voltage to the datalogger VO can be calculated
using the voltage-divider formula (Equation 6.1). This equation
assumes that the input impedance to the Siesta datalogger is so high that it
may be considered an open circuit, especially in relation to the output
resistance of the circuit in Figure 6.5. The sensor
didn't need to accurately detect the amount of pressure being applied, merely
the relative difference in pressure for different situations, so some
deviation was acceptable.
 |
|
|
(6.1) |
 |
|
|
|
The fixed resistance values for R in each of the voltage dividers were
selected by re-arranging Equation 6.1, with VO set to the
maximum required input signal voltage to the datalogger, and RF the
smallest expected sensor resistance. To reduce the risk of damage to the
datalogger, VO was chosen to fall within the datalogger's second
largest gain of 100 millivolts peak-to-peak6.1.
The resistance ranges of the
cm Vermahide sensors built were
measured so appropriate voltage divider R values could be calculated.
Typical value ranges are shown in Table 6.1. The voltage
divider circuit produces the largest signal voltage for the smallest pressure
sensor resistance, and the minimum value of RF (including a safety margin)
under maximum load was taken to be around 15 to 20
.
Using these resistance readings, the voltage divider formula and practical
trials (results shown in Figure 6.7), it was established
that suitable voltage divider biasing resistances were in the range
to
.
As different sensor placements were subject to
different amounts of pressure, and larger R values produce larger signal
voltages for less pressure, it was decided to implement two voltage dividers
with
and two using
.
The
voltage
dividers were used with sensors receiving the most pressure, as is found under
the heel, and the
voltage dividers were used with sensors placed
under the toe. This resulted in all the signals varying over a similar range
in response to their range of applied pressure, which in turn helped produce
simpler and more reliable feature extraction.
Table 6.1:
Vermahide heel pressure sensor resistance measurements
| Applied Load |
Measured resistance |
|
| No load (in air) |
1M |
- | 1.5M |
| Placed in shoe in-sole |
100k |
- | 200k |
| Body weight (double support) |
30k |
- | 40k |
| Body weight (single support on sensor) |
20k |
- | 30k |
|
It was decided that the best pressure sensor positions for collecting useful
ambulation information was under the heel and toe, as these positions play
crucial roles in the gait cycle. The heel is (usually) the point of contact
for a foot-strike, and the toe-off event marks the start of the single-support
phase (See section 2.1.1).
Heel sensors were placed as close the the centre of the Calcaneum (heel bone)
as possible, so that it would receive most pressure during mid stance (tibia
vertical) phase. Two toe sensor positions were investigated; under the base of
the toes (between Metatarsals and Phlanges joints) and under the big toe.
Sensors were either placed under, or attached to, the shoe in-sole using tape.
The in-sole sits between the foot itself and the sole6.2 of the shoe. Although the sensors were not
in direct contact with the foot, the foam in-sole tended to `mold' around the
sensor, discouraging sensors from moving laterally and out of position.
Additionally, this method was found to be more comfortable for the subject.
During trials, it was noticed that the sensor response relied heavily on
placement - if a heel sensor was situated towards the side of the foot, it
would produce smaller signals, which may result in significant gait cycle
events being missed. Although DBNs can compensate for observation errors
through sensor models and conclusions made from other evidence, it was still
preferable to keep sensor readings as accurate as possible.
The sensors and accompanying circuits were designed to produce a peak voltage
around 50 mV for maximum pressure (to suit the datalogger's 100 mV peak-peak
gain setting), and less than 10 mV for no pressure. Table
6.2 shows some typical voltage readings taken from the
sensors under different loads.
Table 6.2:
Vermahide sensor and circuit voltage response
| Applied Load |
Output voltage |
|
| No load (in air) |
0.2m |
- | 3 mV |
| Placed in shoe in-sole |
3m |
- | 6 mV |
| Body weight (double support) |
15m |
- | 30 mV |
| Body weight (single support) |
30m |
- | 50 mV |
|
Some factors influencing sensor response included placement, properties of the
shoe and in-sole, battery voltages, body mass and physiological parameters of
the subject. These factors and the inherent physical differences between
individual sensors meant the signals were not identical. Accordingly, after
sensors were placed, their responses were tested by standing in double
support, then concentrating pressure over each of the sensors in turn. The
results for such a test is shown in Figure 6.7, where he
difference in signals for no sensor load, double support and concentrated
pressure over a sensor can be seen clearly. As long as there were distinct
differences between low and high pressure, the software could be calibrated to
trigger `up' and `down' evidence for the appropriate signal voltages.
Figure 6.7:
Annotated Vermahide pressure transducer signal test
|
|
Some trials were performed with Jean McInerney (see Section
1.4.1), to test the possibility of using electromyograph
(EMG) signals for collecting ambulation-related muscles activity information.
EMG sensors pick up the electrical activity which occurs within muscles when
they move via electrodes coated in special conductive gel placed on the skin
above the muscle of interest. Trials were performed with sensors attached to
calves and and thighs. Initial test results are shown in Figure
6.9.
Figure 6.8:
Compumedics piezoelectric (left) and EMG sensors (right)
|
|
Figure 6.9:
EMG trial signals (top), piezoelectric
accelerometer signals from foot (bottom)
|
|
After some trials, it was decided it was not worthwhile continuing with EMG
sensor trials for a number of reasons. Firstly the EMG sensors are quite
difficult to set up, requiring ground and reference electrodes - usually
attached to the neck or head - as well as those attached over the muscles.
Reference electrodes have to be placed carefully so they can be used to remove
unwanted signals and interference from elsewhere in the body. Secondly, the
EMG signals are characteristically very small in magnitude - in the range of
500 microvolts, and are susceptible to interference from other organs,
external power supplies, crosstalk or capacitance between electrode leads and
lead or sensor movement.
Next: Software
Up: No Title
Previous: The Siesta Datalogger
Daniel J Willis
2000-10-23